| Home | E-Submission | Sitemap | Contact Us |  
top_img
J Radiat Prot > Volume 14(1); 1989 > Article
Journal of Radiation Protection 1989;14(1):66-0.
방사성 중성자선원에 의한 방사선방어측정기의 교정을 위한 표준 중성자 조사장치 연구
최길웅;이경주;황선태;
Standard Neutron Irradiation Facility for Calibration of Radiation Protection Instruments by Radioactive Neutron Sources
Choi, Kil-Oung;Lee, Kyung-Ju;Hwang, Sun-Tae;
ABSTRACT
In routine testing, the radioactive neutron sources are particularly suitable for producing standard. neutron fields. The ISO TC-85 has proposed neutron reference radiation for the calibration of neutron measuring devices used for radiation protection purposes. Radiation laboratory of KSRI has installed a standard irradiation facility using $^{252}Cf$ and $^{241}Am-Be$ sources for calibrating personal dosimeters according to the recommendations given in ISO TC-85. In this study, correction factors for calibration related to neutron scattering and anisotropy are obtained by experiments with commercial rem meter for demonstration purposes.