| Home | E-Submission | Sitemap | Contact Us |  
top_img
J Radiat Prot > Volume 10(1); 1985 > Article
Journal of Radiation Protection 1985;10(1):14-0.
형광 X 선을 이용한 100 KeV 이하의 에너지 영역에서의 단색 Photon 선원개발에 관한 연구
이연명;이건재;하석호;황선태;이경주;
Development of Monoenergetic Photon Source in the Energe Range below 100 keV by the X-ray Fluorescence Method
Lee, Youn-Myoung;Lee, Kun-Jai;Hah, Suck-Ho;Hwang, Sun-Tae;Lee, Kyung-Ju;
ABSTRACT
The development of mono energetic photon sources using $K_{alpha}$ fluorescence X-ray of pure material was carried out in the energy range below 100 keV. The monoenergetic photons are very useful in the calibration of the radiation measuring instruments and can be produced as the $K_{alpha}$ fluorescence X-ray by irradiating the bremsstrahlung to the thin pure metal foils called ‘radiators’. In this experiment, several radiators such as $_{47}Ag,;_{50}Sn,;_{68}Er,;_{70}Yb,;and;_{82}Pb$ provide the wide monoenergetic photon energy ranging from 20 keV to 80 keV. By the spectrometry with HpGe LEPS, spectral purity factors which measure the monochrometicity for the $K_{alpha}$ fluorescence X-ray, were determined as $0.64{sim}0.94$. Dosimetry for the purpose of the determination of the exposure rate with a 600cc thin window ionization chamber, which was calibrated by the standard free-air ionization chamber, was performed. Exposure rates ranging $8.3{sim}232.5mR/h$ was obtained according to the $K_{alpha}$ fluorescence X-ray energy for each radiator.
TOOLS
PDF Links  PDF Links
Full text via DOI  Full text via DOI
Download Citation  Download Citation
Share:      
METRICS
1,112
View
1
Download
Determination of Personnel Exposures in the Lower Energy Ranges of X-Ray by Photographic Dosimeter  1986 ;11(1)
Editorial Office
#319, Hanyang Institute of Technology Bldg., 222 Wangsimni-ro, Seongdong-gu,Seoul, Republic of Korea
Tel: +82-2-2297-9775   Fax: +82-2-2297-9776
Email: managing.editor@jrpr.org
About |  Browse Articles |  Current Issue |  For Authors and Reviewers
Copyright © by Korean Association for Radiation Protection. Developed in M2PI