| Home | E-Submission | Sitemap | Contact Us |  
top_img
J Radiat Prot > Volume 13(2); 1988 > Article
Journal of Radiation Protection 1988;13(2):29-0.
휴대형 고순도 게르마늄검출기에 대한 스펙트럼-조사선량율 변환연산자의 결정
곽상수;박종묵;노성기;
Determination of Spectrum-Exposure Rate Conversion Factor for a Portable High Purity Germanium Detector
Kwak, Sang-Soo;Park, Chong-Mook;Ro, Seung-Gy;
ABSTRACT
A spectrum-exposure rate conversion operator G(E) for a portable HPGe detector used for field environmental radiation survey was theoretically developed on the basis of a space distribution function of gamma flux emitted from a disk source and an areal efficiency of the detector. The radiation exposure rates measured using this G(E) and the portable HPGe. detector connected to a portable multichannel analyzer were compared with those measured by a 3' ${phi};{times}$3' NaI(Tl) scintillation detector with the reported G(E) and a pressurized ionization chamber. A comparison of the three results showed that the result obtained using the HPGe detector was lower than those determined using the NaI(Tl) detector and ionization chamber by 17% to 29%, The difference obtained is close to that reported in literature. The method developed here can be easily applicable to obtain a G(E) factor suitable to any detector for detecting the exposure rate of environmental gamma radiation, since the spectrum-exposure rate conversion operator can be calculated by a hand calculator.
Editorial Office
#319, Hanyang Institute of Technology Bldg., 222 Wangsimni-ro, Seongdong-gu,Seoul, Republic of Korea
Tel: +82-2-2297-9775   Fax: +82-2-2297-9776
Email: jrpr.editorial@gmail.com
About |  Browse Articles |  Current Issue |  For Authors and Reviewers
Copyright © by Korean Association for Radiation Protection. Developed in M2PI