ABSTRACT
The optimum etching condition of allyl diglycol carbonate (CR-39) for detecting the 60 MeV N-ions was determined as $70^{circ}C$, 20% NaOH for 130min, by considering the variations of track density and diameter. Under these conditions, the maximum detectable track density was $1.7{times}10^7tr/cm^2$. Track densities were linearly increased with increase of the total charge of the incident 60 MeV N-ions. By considering the scattring of N-ions as the Rutherford elastic scattering of point source, the measured relative probability was well consistent with the calculated value. The detection efficiency of CR-39 was in the range of 54-41% for 60 MeV N-ions of 12 nC-100 nC.