To know which parameters were acceptable for achieving lowest radiation exposure to the patients and highest image quality at the diagnostic X-ray radiography, we measured the patient radiation dose and image quality in transmitted PACS (Picture Archiving and Communication System) at variable combinations of the added filters. As a result, the Dose Area Product (DAP: $mGy{cdot}cm^2$) and Entrance Surface Doses (ESDs: $mGy$) was lowest at 1 mmAl + 0.2 mmCu and highest at 0 mmAl. The histogram of the image quality by transmitted PACS was not significantly different at variable combinations of exposure parameters on the MATLAB. In conclusion, this study can be helpful for expecting radiation dose-exposure and control exposure parameters for the diagnostic X-ray radiography.