| Home | E-Submission | Sitemap | Contact Us |  
top_img
J. Radiat. Prot. Res > Volume 48(3); 2023 > Article
Son, Lee, Choi, Park, and Kim: Impact of 0.35 T Magnetic Field on Dose Calculation for Non-small Cell Lung Cancer Stereotactic Radiotherapy Plans

Abstract

Background

We investigated the impact of 0.35 T magnetic field on dose calculation for non-small cell lung cancer (NSCLC) stereotactic ablative radiotherapy (SABR) in the ViewRay system (ViewRay Inc.), which features a simultaneous use of magnetic resonance imaging (MRI) to guide radiotherapy for an improved targeting of tumors.

Materials and Methods

Here, we present a comprehensive analysis of the effects induced by the 0.35 T magnetic field on various characteristics of SABR plans including the plan qualities and dose calculation for the planning target volume, organs at risk, and outer/inner shells. Therefore, two SABR plans were set up, one with a 0.35 T magnetic field applied during radiotherapy and another in the absence of the field. The dosimetric parameters were calculated in both cases, and the plan quality indices were evaluated using a Monte Carlo algorithm based on a treatment planning system.

Results and Discussion

Our findings showed no significant impact on dose calculation under the 0.35 T magnetic field for all analyzed parameters. Nonetheless, a significant enhancement in the dose was calculated on the skin surrounding the tumor when the 0.35 T magnetic field was applied during the radiotherapy. This was attributed to the electron return effect, which results from the deviation of the electrons ejected from tissues upon radiation due to Lorentz forces. These returned electrons re-enter the tissues, causing a local dose increase in the calculated dose.

Conclusion

The present study highlights the impact of the 0.35 T magnetic field used for MRI in the ViewRay system for NSCLC SABR treatment, especially on the skin surrounding the tumors.

Introduction

Non-small cell lung cancer (NSCLC) is a major cause of cancer-related mortality worldwide, with more than 1,000,000 related deaths occurring each year [13]. Although surgery is the most representative treatment for NSCLC cases with high mortality rates, it does not constitute a default option for a significant percentage of patients owing to various reasons. In fact, approximately 30% of early-stage NSCLC patients do not have surgery, with 65% of elderly patients older than 75 years [4]. Additionally, patients with comorbidities, which may increase surgical risk, generally do not undergo surgery [57]. Nonetheless, a decision not to undergo surgery can also be due to a patient’s perception of prognosis and racial factors. Thus, for patients excluding the surgery option, Stereotactic ablative radiotherapy (SABR) is the alternative route to conventional fractionated radiotherapy [810]. SABR is a radiotherapy method with extremely high ablation doses, usually provided in three to eight treatment fractions within 2 weeks, to minimize the toxicity risk induced to normal cells. Many studies have demonstrated that SABR treatment for NSCLC patients achieves better local control, survival rates, and less toxicity, compared to conventional radiotherapy [5, 11, 12]. Nonetheless, the SABR treatment requires a set of logistics including immobilization to prevent patient movements during radiation treatment, treatment plans, and actions to describe tumor movements during radiation delivery, dose distribution for tumors, and steep dose slopes from tumors to surrounding normal tissue [13, 14]. SABR is a method of high-precision radiation therapy with a relatively larger amount of irradiation per fraction than conventional fractionated radiotherapy, which imposes an essential need for highly accurate treatment. The ViewRay system (ViewRay Inc.) provides means to deal with most of the SABR treatment requirements. Magnetic resonance imaging (MRI) is performed in real-time, with a 0.35 T magnetic field, during radiation treatment [1517]. This enables accurate doses to be delivered to the targeted tumor. However, the low magnetic strength used in ViewRay infers poor image quality. Most importantly, the magnetic field during treatment affects dose calculations, causing dose enhancement or reduction. Raaijmakers et al. [18] reported a reduced build-up distance induced by the magnetic field because it creates an electron trajectory between collisions in the form of curves. In addition, a dose enhancement usually arises due to the electron return effect (ERE) at tissue-air interfaces. Electrons leave the tissue at a certain rate and then deviate from the path by the magnetic field, curve in the air and enter the tissue again, causing a local dose enhancement. Raaijmakers et al. [19] found that ERE increased the dose enhancement by 40% at the beam exit region. This phenomenon affects the dose depositions in the clinical planning stage [19]. Esmaeeli et al. [20] reported the effects of a magnetic field (1.5 or 3.0 T) on dose distribution in breast-cancer plans. Besides, Kim et al. [21] reported the impact of a low magnetic field on dose calculation for partial breast irradiation. In particular, observations on the outside of the body confirmed maximum dose (Dmax) showed a 5.0±10.5 Gy dose difference. Moreover, Son et al. [22] evaluated the influence of the magnetic field in the prostate SABR plan. It was reported that Dmax differs by 1.28±1.08 Gy on the outside of the body, confirming that the magnetic field has an influence at the tissue-air interface, inducing alterations to the dose calculation. The tumors in patients with NSCLC are surrounded by more air, compared to the tumors in patients with partial breast or prostate cancer. Therefore, in this study, we evaluated the effect of a 0.35 T magnetic field used in ViewRay on dose deposition. Clinically relevant dosimetric parameters and plan qualities were analyzed by comparing the presence/absence of a 0.35 T magnetic field for NSCLC SABR plans.

Materials and Methods

1. Patient Selection

Twenty patients who have undergone SABR treatment for NSCLC between January 2014 and February 2022 were selected using ViewRay, retrospectively.

2. Treatment System (ViewRay)

The ViewRay system consists of a radiation therapy technique using three 60Co sources, accompanied by MRI with a 0.35 T magnetic field. The system was equipped with three depleted 60Co heads on a ring-shaped gantry. Each multi-leaf collimator (MLC) used in this system is composed of two leaf banks equipped with a total of 60 equally mounted double-focus leaves (30 on each leaf bank). The leaf width of the MLC was 1.05 cm at the plane of the isocenter at a source to axis distance of 105 cm. Each leaf had a 0.3 mm thickness of tongue and groove on the leading edge and a 0.5 mm thickness on the adjacent sides. The interval gap of MLCs is 0.5 cm with a maximum aperture size of 27.3 cm×27.3 cm at the isocenter.

3. SABR Plans for Lung Cancer

The treatment plans without main magnetic field (B0)=0 T treatment plans that are being compared with the B0=0.35 T plans were not created from scratch. Instead, the B0=0 T plans were recalculated (but not re-optimized) with B0= 0.35 T plans to determine the extent to which the 0.35 T magnetic field affects dose depositions. The volume of planning target volumes (PTV) has a range of 3.0 to 41.0 cm3, with an average volume of 15.6 cm3. The isocenter was laterally fixed in order to prevent collision between the bore of the 60Co system and the patient’s arms. The dose was calculated with an radiofrequency coil using a 3 mm calculation grid size. The dose calculation of the 60Co system is achieved based on the Monte Carlo (MC) calculation algorithm developed by the manufacturer (ViewRay Inc.). The dose volume histogram (DVH) was analyzed for dosimetric parameters obtained from the ViewRay system.

4. Statistics Analysis

We used the IBM SPSS version 23.0 statistical software (IBM Co.) for the analysis. We used the paired t-test to evaluate the difference in dose deposition with and without a 0.35 T magnetic field. The paired t-test compares the means of two results taken from the related targets. The results are presented as the average±standard deviation. Paired t-tests and paired-sample comparisons were calculated.

5. Plan Evaluation

We used the DVHs to assess the impact of the 0.35 T magnetic field. The corresponding results were analyzed for the presence/absence of a low magnetic field, respectively. For the PTV, we compared the results of dose for mean dose (Dmean), minimum dose (Dmin), Dmax, the dose delivered to the 5%, 90%, 95%, and 100% of the volume (D5%, D90%, D95%, and D100%, respectively). For organs at risks (OARs), the dosimetric parameters were compared with Dmax for the spinal cord; Dmax for the heart; Dmax for bronchi; Dmax and the volume receiving doses equal to or greater than 40 Gy (V40Gy) for ribs; Dmax for skin; Dmean, the volume receiving doses equal to or greater than 10 Gy (V10Gy), the volume receiving doses equal to or greater than 5 Gy (V5Gy), and the dose delivered to the volume 1,000 cm3 (D1000cm3) for the contralateral lung; and Dmean, the volume delivered to the 20 Gy (V20Gy), V10Gy, V5Gy, and D1000cc for the ipsilateral lung. We also analyzed Dmean, Dmin, and Dmax for the inner and outer shells (±3 mm from body surface) in order to evaluate the impact of the magnetic field in the patient surface region. Furthermore, the change in the plan’s quality was quantitatively evaluated based on the change in dosimetric parameters. The homogeneity index (HI), conformity index (CI), and target coverage were analyzed to evaluate the plan’s qualities. The HI, which quantifies the dose homogeneity in the PTV, was calculated as the ratio of D5% to D95% [23]. The CI, which indicates the relationship between isodose distribution and PTV, was calculated as the ratio of the prescription dose delivered volume to the PTV volumes [24]. All plans have a target coverage of 95% for the target volume, receiving 100% of the prescription dose.

Results and Discussion

Fig. 1 shows the comparison results of the dose deposition in the presence/absence of a 0.35 T magnetic field.

1. PTV Dose Distribution Difference

The calculated dose deposition for PTV was obtained with/ without a 0.35 T magnetic field. The analyzed doses were Dmean, Dmin, Dmax, D5%, D90%, D95%, and D100%, as listed in Table 1. The average differences in dose depositions between the cases with/without the magnetic field were determined at Dmean, Dmin, and Dmax as 0.21±0.18 Gy (p=0.36), 0.48±0.34 Gy (p= 0.38), and 0.58±0.43 Gy (p=0.38), respectively. At D5%, D90%, D95%, and D100%, the average differences in dose deposition were calculated to be 0.31±0.30 Gy (p=0.35), 0.27±0.28 Gy (p=0.38), 0.28±0.29 Gy (p=0.39), and 0.48±0.34 Gy (p=0.38), respectively.
These results show very small differences between the dose calculations (but only one optimization) in the presence and absence of a 0.35 T magnetic field. This clearly indicates that the 0.35 T magnetic field, used for MRI in the ViewRay system, has no significant effect on the dose depositions for the PTV.

2. Plan Qualities Difference

Table 2 summarizes the plan qualities of the NSCLC SABR plans in the presence/absence of a low magnetic field. The results were analyzed for HI, CI, and target coverage. In the presence of a 0.35 T magnetic field, the following results were obtained for these parameters: HI=1.01±0.05; CI=1.25± 0.016; and target coverage=0.95±0.02. When the calculation was performed without the magnetic field, the following results were obtained: HI=1.10±0.05; CI=1.28±0.17; and target coverage=0.94±0.05. For target coverage, the p-value was determined at 0.18 for the presence/absence of a magnetic field. Therefore, the comparison of these indicators clearly indicates that the plan quality is not affected by the presence and absence of the 0.35 T magnetic field.

3. Organs at Risk Dose Distribution Difference

Table 3 presents the average difference for the OARs including the spinal cord, heart, bronchi, ribs, skin, contralateral lung, and ipsilateral lung. The average differences in Dmax for the spinal cord, heart, and bronchi, were determined as 0.13±0.12 Gy (p=0.50), 0.21±0.17 Gy (p=0.50), and 0.05± 0.08 Gy (p=0.50), respectively. For the ribs, the differences in Dmax and V40Gy were 0.20±0.28 Gy (p=0.40) and 0.15±0.45% (p=0.45), respectively. For the opposite lung where the PTV was located, Dmean and D1000cc yielded average differences of 0.04±0.13 Gy (p=0.45) and 0.01±0.01 Gy (p=0.42), respectively. For V10Gy and V5Gy, the average differences were 0.01± 0.02% (p=0.42) and 0.06±0.07% (p=0.45), respectively. For the ipsilateral lung, the average differences for Dmean, D1000cc, V20Gy, V10Gy, and V5Gy were 0.72±2.30 Gy (p=0.38), 0.02±0.02 Gy (p=0.46), 0.17±0.29% (p=0.49), 0.13±0.15% (p=0.50), and 0.22±0.46% (p=0.49), respectively. The very small differences obtained between the values in the presence and absence of the low magnetic field indicate that these conditions do not affect the dosimetric parameters for various OARs.

4. Dose Distribution Difference around the Skin

The dose difference around the skin in the presence/absence of a low magnetic field is shown in Table 4. For the inner shell, the dose differences in terms of Dmean, Dmin, and Dmax were 0.03±0.02 Gy (p=0.47), 0.04±0.13 Gy (p=0.15), and 0.47±0.33 Gy (p=0.41), respectively. No statistically significant differences were observed for the inner shell. However, for the outer shell, the differences in terms of Dmean, Dmin, and Dmax were 0.08±0.06 Gy (p=0.45), 0.02±0.04 Gy (p=0.15), and 3.58±2.60 Gy (p=0.05), respectively. This result means that the Dmax of the outer shell is significantly affected by the presence of the 0.35 T magnetic field.
The impact on the dose deposition of the 0.35 T magnetic field, applied in ViewRay to guide radiation therapy, was analyzed for SABR plans for NSCLC in terms of the variation in dose distributions for PTV, OARs, and outer/inner shells. The effect on plan quality indices (i.e., HI, CI, and target coverage) was also analyzed. For this, two NSCLC SABR plans were designed: one original plan with a 0.35 T magnetic field and another in the absence of the magnetic field. Although no significant differences were observed in PTV, plan qualities, and OARs, results showed a significant increase in doses outside the body. This indicates that the ERE induces variations in the dose deposition around the skin in SABR plans for NSCLC.
The MC algorithm based on a treatment planning system (TPS) was employed to evaluate the treatment plans. To use high performance for the optimized simulation of photons, the TPS system applied a variance reduction technique. Dose calculation based on the MC algorithm in the presence/absence of a magnetic field provides geometric information of secondary electrons generated by photons [25]. These electrons can be bent by Lorentz forces and thus have arc-shaped trajectories in water or tissue, causing physical changes such as reduced accumulation, increased doses, and penumbra [26]. In this case, the deviation in dose distribution might have increased in the body. However, Esmaeeli et al. [20] have shown that a dose distribution with an increase or decrease of 7% or 12%, respectively, might occur near the surface. They argued that the increase or decrease in dose to the PTV, internal and contralateral tissues could be reduced because of the magnetic field. In contrast, we have confirmed in our previous studies that low magnetic fields increase the dose, especially near the body surfaces [21, 25]. In the present study, our findings demonstrate that, in the PTV and internal tissues, the dose distribution with magnetic fields did not change compared to the case where condition without the magnetic field was applied for breast and liver cancers [21, 25]. Nonetheless, we report on a slight increase in dose due to the ERE outside the body. In addition, electrons may be more likely to build-up again, especially in air.
These results indicate that further studies are needed on the impact of low magnetic fields for different treatment techniques and sites. In addition, Unity (Elekta) has been developed to allow 6 MV planarization filter free beams to be irradiated to targets, while simultaneously obtaining images using 1.5 T magnetic fields [27]. In the case of a 1.5 T magnetic field, the differences in dose deposition for the PTV, OARs, and skin are greater than those caused by the impact on the 0.35 T magnetic field [28].

Conclusion

This study shows that a 0.35 T magnetic field has an impact on the dose depositions in NSCLC SABR plans. In particular, there was a significant increase in the outer shell dose under the 0.35 T magnetic field due to scatter radiation or the secondary electrons. No significant differences of dose depositions were identified in the PTV, plan quality, and OARs.

Notes

Conflict of Interest

No potential conflict of interest relevant to this article was reported.

Ethical Statement

This study was approved by the Institutional Review Board of Seoul National University Hospital (approval number: H-2008-200-1155). Written informed consent by the patients was waived due to a retrospective nature of our study.

Author Contribution

Conceptualization: Kim J. Methodology: Choi CH, Park JM. Formal analysis: Lee SY. Funding acquisition: Kim J. Project administration: Kim J. Visualization: Son J. Writing - original draft: Son J. Writing - review & editing: Son J. Approval of final manuscript: all authors.

Acknowledgements

This research was supported by the National Research Foundation of Korea (NRF) grant funded by the Korean government (Ministry of Science and ICT) (Grant No. 2021M2E 8A1049195) and (Grant No. 2019K1A3A1A80113183).

References

1. Goldstraw P, Ball D, Jett JR, Le Chevalier T, Lim E, Nicholson AG, et al. Non-small-cell lung cancer. Lancet. 2011;378(9804):1727-1740.
crossref pmid
2. Ettinger DS, Akerley W, Bepler G, Blum MG, Chang A, Cheney RT, et al. Non-small cell lung cancer. J Natl Compr Canc Netw. 2010;8(7):740-801.
crossref pmid
3. Brown S, Banfill K, Aznar MC, Whitehurst P, Faivre Finn C. The evolving role of radiotherapy in non-small cell lung cancer. Br J Radiol. 2019;92(1104):20190524.
crossref pmid pmc
4. Palma D, Visser O, Lagerwaard FJ, Belderbos J, Slotman BJ, Senan S. Impact of introducing stereotactic lung radiotherapy for elderly patients with stage I non-small-cell lung cancer: a population-based time-trend analysis. J Clin Oncol. 2010;28(35):5153-5159.
crossref pmid
5. Lagerwaard FJ, Verstegen NE, Haasbeek CJ, Slotman BJ, Paul MA, Smit EF, et al. Outcomes of stereotactic ablative radiotherapy in patients with potentially operable stage I non-small cell lung cancer. Int J Radiat Oncol Biol Phys. 2012;83(1):348-353.
crossref pmid
6. Le Chevalier T, Arriagada R, Quoix E, Ruffie P, Martin M, Tarayre M, et al. Radiotherapy alone versus combined chemotherapy and radiotherapy in nonresectable non-small-cell lung cancer: first analysis of a randomized trial in 353 patients. J Natl Cancer Inst. 1991;83(6):417-423.
crossref pmid
7. Finlayson E, Fan Z, Birkmeyer JD. Outcomes in octogenarians undergoing high-risk cancer operation: a national study. J Am Coll Surg. 2007;205(6):729-734.
crossref pmid
8. Solda F, Lodge M, Ashley S, Whitington A, Goldstraw P, Brada M. Stereotactic radiotherapy (SABR) for the treatment of primary non-small cell lung cancer: systematic review and comparison with a surgical cohort. Radiother Oncol. 2013;109(1):1-7.
crossref pmid
9. Cao C, D’Amico T, Demmy T, Dunning J, Gossot D, Hansen H, et al. Surgery versus SABR for resectable non-small-cell lung cancer. Lancet Oncol. 2015;16(8):e370-e371.
crossref
10. Iyengar P, Westover K, Timmerman RD. Stereotactic ablative radiotherapy (SABR) for non-small cell lung cancer. Semin Respir Crit Care Med. 2013;34(6):845-854.
crossref pmid
11. Verstegen NE, Lagerwaard FJ, Haasbeek CJ, Slotman BJ, Senan S. Outcomes of stereotactic ablative radiotherapy following a clinical diagnosis of stage I NSCLC: comparison with a contemporaneous cohort with pathologically proven disease. Radiother Oncol. 2011;101(2):250-254.
crossref
12. Abel S, Hasan S, White R, Schumacher L, Finley G, Colonias A, et al. Stereotactic ablative radiotherapy (SABR) in early stage non-small cell lung cancer: comparing survival outcomes in adenocarcinoma and squamous cell carcinoma. Lung Cancer. 2019;128:127-133.
crossref pmid
13. Vlaskou Badra E, Baumgartl M, Fabiano S, Jongen A, Guckenberger M. Stereotactic radiotherapy for early stage non-small cell lung cancer: current standards and ongoing research. Transl Lung Cancer Res. 2021;10(4):1930-1949.
crossref pmid pmc
14. Goldsmith C, Gaya A. Stereotactic ablative body radiotherapy (SABR) for primary and secondary lung tumours. Cancer Imaging. 2012;12(2):351-360.
crossref pmid pmc
15. Park JM, Park SY, Kim HJ, Wu HG, Carlson J, Kim JI. A comparative planning study for lung SABR between tri-Co-60 magnetic resonance image guided radiation therapy system and volumetric modulated arc therapy. Radiother Oncol. 2016;120(2):279-285.
crossref pmid
16. Choi CH, Park SY, Kim JI, Kim JH, Kim K, Carlson J, et al. Quality of tri-Co-60 MR-IGRT treatment plans in comparison with VMAT treatment plans for spine SABR. Br J Radiol. 2017;90(1070):20160652.
crossref pmid pmc
17. Luterstein E, Cao M, Lamb JM, Raldow A, Low D, Steinberg ML, et al. Clinical outcomes using magnetic resonance-guided stereotactic body radiation therapy in patients with locally advanced cholangiocarcinoma. Adv Radiat Oncol. 2019;5(2):189-195.
crossref pmid pmc
18. Raaijmakers AJ, Raaymakers BW, Lagendijk JJ. Integrating a MRI scanner with a 6 MV radiotherapy accelerator: dose increase at tissue-air interfaces in a lateral magnetic field due to returning electrons. Phys Med Biol. 2005;50(7):1363-1376.
crossref pmid
19. Raaijmakers AJ, Raaymakers BW, Lagendijk JJ. Magnetic-field-induced dose effects in MR-guided radiotherapy systems: dependence on the magnetic field strength. Phys Med Biol. 2008;53(4):909-923.
crossref pmid
20. Esmaeeli AD, Pouladian M, Monfared AS, Mahdavi SR, Moslemi D. Effect of uniform magnetic field on dose distribution in the breast radiotherapy. Int J Radiat Res. 2014;12(2):151-160.

21. Kim JI, Park SY, Lee YH, Shin KH, Wu HG, Park JM. Effect of low magnetic field on dose distribution in the partial-breast irradiation. Prog Med Phys. 2015;26(4):208-214.
crossref
22. Son J, An HJ, Choi CH, Chie EK, Kim JH, Park JM, et al. Assessment of dose distributions according to low magnetic field effect for prostate SABR. J Radiat Prot Res. 2019;44(1):26-31.
crossref
23. Kataria T, Sharma K, Subramani V, Karrthick KP, Bisht SS. Homogeneity index: an objective tool for assessment of conformal radiation treatments. J Med Phys. 2012;37(4):207-213.
crossref pmid pmc
24. Han G, Liu D, Gan H, Denniston KA, Li S, Tan W, et al. Evaluation of the dosimetric feasibility of hippocampal sparing intensity-modulated radiotherapy in patients with locally advanced nasopharyngeal carcinoma. PLoS One. 2014;9(2):e90007.
crossref pmid pmc
25. Son J, Chun M, An HJ, Kang SH, Chie EK, Yoon J, et al. Effect of low magnetic field on dose distribution in the SABR plans for liver cancer. Prog Med Phys. 2018;29(2):47-52.
crossref
26. Oborn BM, Metcalfe PE, Butson MJ, Rosenfeld AB. Monte Carlo characterization of skin doses in 6 MV transverse field MRI-linac systems: effect of field size, surface orientation, magnetic field strength, and exit bolus. Med Phys. 2010;37(10):5208-5217.
crossref pmid
27. Roberts DA, Sandin C, Vesanen PT, Lee H, Hanson IM, Nill S, et al. Machine QA for the Elekta unity system: a report from the Elekta MR-linac consortium. Med Phys. 2021;48(5):e67-e85.
crossref pmid pmc
28. Kubota T, Araki F, Ohno T. Comparison of dose distributions between transverse magnetic fields of 0.35 T and 1.5 T for radiotherapy in lung tumor using Monte Carlo calculation. Med Dosim. 2020;45(2):179-185.
crossref pmid

Fig. 1
An example of dose calculation for non-small cell lung cancer stereotactic radiotherapy plan: (A) with a magnetic field (B0=0.35 T), (B) without a magnetic field.
jrpr-2022-00150f1.jpg
Table 1
Planning Target Volume Dosimetric Parameters with/without Magnetic Field
Parameter With magnetic field Without magnetic field Difference value p-value
Dmean (Gy) 56.50±5.41 56.62±5.28 0.21±0.18 0.356
Dmin (Gy) 51.01±5.37 51.02±5.63 0.48±0.34 0.380
Dmax (Gy) 60.09±5.67 60.13±5.12 0.58±0.43 0.381
D5% 58.99±5.48 59.16±5.10 0.31±0.30 0.354
D90% 54.38±5.51 54.37±5.57 0.27±0.28 0.377
D95% 53.79±3.52 53.73±5.60 0.28±0.29 0.387
D100% 51.01±5.37 51.02±5.63 0.48±0.34 0.380

Values are presented as mean±standard deviation.

Dmean, mean dose; Dmin, minimum dose; Dmax, maximum dose; D5%, dose delivered to the 5% of the volume; D90%, dose delivered to the 90% of the volume; D95%, dose delivered to the 95% of the volume; D100%, dose delivered to the 100% of the volume.

Table 2
Plan Quality Indices with/without Magnetic Field
Parameter With magnetic field Without magnetic field p-value
Homogeneity index 1.099±0.052 1.100±0.052 0.479
Conformity index 1.253±0.155 1.282±0.169 0.289
Target coverage 0.946±0.018 0.935±0.049 0.178

Values are presented as mean±standard deviation.

Table 3
Organs at Risk Dosimetric Parameters with/without Magnetic Field
Parameter With magnetic field Without magnetic field Difference value p-value
Spinal cord
 Dmax (Gy) 11.10±4.26 11.10±4.33 0.13±0.12 0.499

Heart
 Dmax (Gy) 19.13±11.09 19.15±11.08 0.21±0.17 0.498

Bronchi
 Dmax (Gy) 7.03±14.44 6.98±14.44 0.05±0.08 0.497

Ribs
 Dmax (Gy) 53.25±6.57 52.41±6.20 0.20±0.28 0.402
 V40Gy (%) 3.41±3.05 3.60±3.07 0.15±0.45 0.451

Skin
 Dmax (Gy) 60.09±5.67 59.92±5.42 0.59±0.42 0.462

Contralateral lung
 Dmean (Gy) 1.16±0.59 1.13±0.63 0.04±0.13 0.445
 V10Gy (%) 2.02±2.56 2.37±2.61 0.01±0.02 0.420
 V5Gy (%) 5.21±5.72 4.94±5.74 0.06±0.07 0.447
 D1000cm3 (Gy) 0.27±0.11 0.28±0.10 0.01±0.01 0.418

Ipsilateral lung
 Dmean (Gy) 6.36±2.86 6.69±3.35 0.72±2.30 0.375
 V20Gy (%) 11.58±6.97 11.63±6.92 0.17±0.29 0.493
 V10Gy (%) 22.09±10.47 22.07±10.52 0.13±0.15 0.498
 V5Gy (%) 30.95±12.76 30.80±12.71 0.22±0.46 0.485
 D1000cm3 (Gy) 0.69±0.52 0.67±0.52 0.02±0.02 0.462

Values are presented as mean±standard deviation.

Dmax, maximum dose; V40Gy, volume receiving doses equal to or greater than 40 Gy; V10Gy, volume receiving doses equal to or greater than 10 Gy; V5Gy, volume receiving doses equal to or greater than 5 Gy; D1000cm3, dose delivered to the volume 1,000 cm3; Dmean, mean dose; V20Gy, volume receiving doses equal to or greater than 20 Gy.

Table 4
Inner/Outer Shell Dosimetric Parameters with/without Magnetic Field
Parameter With magnetic field Without magnetic field Difference value p-value
Inner shell
 Dmean (Gy) 1.11±0.58 1.13±0.58 0.03±0.02 0.470
 Dmin (Gy) 0.02±0.02 0.06±0.14 0.04±0.13 0.154
 Dmax (Gy) 30.79±4.92 31.15±5.00 0.47±0.33 0.408

Outer shell
 Dmean (Gy) 0.86±0.42 0.87±0.40 0.08±0.06 0.452
 Dmin (Gy) 0.02±0.02 0.03±0.05 0.02±0.04 0.152
 Dmax (Gy) 18.74±6.05 21.20±5.49 3.58±2.60 0.047

Values are presented as mean±standard deviation.

Dmean, mean dose; Dmin, minimum dose; Dmax, maximum dose.

TOOLS
PDF Links  PDF Links
PubReader  PubReader
ePub Link  ePub Link
XML Download  XML Download
Full text via DOI  Full text via DOI
Download Citation  Download Citation
  Print
Share:      
METRICS
0
Crossref
0
Scopus 
1,585
View
147
Download
Editorial Office
#319, Hanyang Institute of Technology Bldg., 222 Wangsimni-ro, Seongdong-gu,Seoul, Republic of Korea
Tel: +82-2-2297-9775   Fax: +82-2-2297-9776
Email: managing.editor@jrpr.org
About |  Browse Articles |  Current Issue |  For Authors and Reviewers
Copyright © by Korean Association for Radiation Protection. Developed in M2PI