ABSTRACT
In order to determine the $^{56}Mn;{gamma}$-detection efficiency of a $MnSO_4$ bath system, it is essential to do the absolute activity measurement of $^{56}Mn$ solution. For the fabrication of $^{56}Mn$ samples, a 13.718 mg of $^{56}Mn$ metal flake with 99.99% purity was irradiated for 12 minutes at the thermal neutron field of about $10^{13}n/cm^2s$ of flux density. The neutron activated $^{56}Mn$ metal sample was dissolved in 50 ml of 0.1 N-HCl solution. The $^{56}Mn$ samples were fabricated by using the dissolved stock solution and the activity of each of them was measured by the $4{pi}{beta}-{gamma}$ coincidence counting technique. The obtained result was 408.070 kBq/mg with total uncertainty of 0.366% at reference date, 0 h on October 15, 1987.