| Home | E-Submission | Sitemap | Contact Us |  
top_img
J Radiat Prot > Volume 38(3); 2013 > Article
Journal of Radiation Protection 2013;38(3):132-0.
doi: https://doi.org/10.14407/jrp.2013.38.3.132
저온검출기를 이용한 에너지 고 분해능 알파분광 구현
김민성;이상훈;윤원식;장용식;이상준;김용함;이민규;
High Energy Resolution Alpha Spectrometer Using a Cryogenic Detector
Kim, M.S.;Lee, S.H.;Yoon, W.S.;Jang, Y.S.;Lee, S.J.;Kim, Y.H.;Lee, M.K.;
ABSTRACT
Cryogenic particle detectors have recently been adopted in radiation detection and measurement because of their high energy resolution. Many of these detectors have demonstrated energy resolutions better than the theoretical limit of semiconductor detectors. We report the development of alpha spectrometer using a micro-fabricated magnetic calorimeter coupled to a large-area particle absorber. A piece of gold foil of $2{times}2{times}0.05mm^3$ was glued to the paramagnetic temperature sensor made of sputtered Au:Er film to serve as an absorber for incident alpha particles. We performed experiments with 241Am source at cryogen free adiabatic demagnetization refrigerator (CF-ADR). A high energy resolution of 6.8 keV in FWHM was obtained for 5.5 MeV alpha particles.